Three dimensional investigation of in vivo anterior eye segments by swept-source optical coherence tomography with ready-for-shipping scanning light source

Three dimensional investigation of in vivo anterior eye segments by swept-source optical coherence tomography with ready-for-shipping scanning light source


概要

Abstract: A two- and three- dimensional swept source optical coherence tomography (SS-OCT) system which uses a ready-to-ship scanning light source is demonstrated. The light source has the center wavelength of 1.31 μm, the -3 dB wavelength range of 110 nm, the scanning rate of 20 KHz and high linearity of frequency scanning. A simple calibration method using a fringe analysis technique for spectral rescaling is presented. This SS-OCT is capable of realtime display of two-dimensional OCT, and can take three-dimensional OCT with the measurement time of 2 s. In vivo human anterior eye segments are investigated both two- and three- dimensionally. The system sensitivity is experimentally determined as 113 dB.



Keywords

OCT, Optical Coherence Tomography, wavelength-swept,

著者

Yoshiaki Yasuno, Violeta Dimitrova Madjarova, Shuichi Makita, Masahiro Akiba, Atsushi Morosawa, Changho Chong, Toru Sakai, Kin-Pui Chan, Masahide Itoh and Toyohiko Yatagai,

発表日

2006

PDF