Think Precise

Multi-channel Optical Power Meter for Swept Test System

MPM-210H
MPM-211, 212, 213, 215

The perfect tool for very fast simultaneous measurement of multi-port optical devices

  • High Speed Measurement
    (Up to 1 million logging sampling points per port)
  • Up to 20-Port Measurement
  • Power Meter Module (MPM-211 / 212 / 215)
  • Current Meter Module (MPM-213)


MPM-210
MPM-210H/MPM-211, 212, 213, 215

Overview

Our MPM-210H is ideal for IL, WDL and PDL measurement of multi-port optical components including Dense Wavelength Division Multiplexing (DWDM), Arrayed Waveguide Gratings(AWG) , Wavelength Selective Switches (WSS) and more. When combined with our TSL-Series laser equipped with a power monitor output, the MPM-210H allows the user to complete real-time high precision IL measurements.




Features & Charateristics

  • High speed measurement
    Up to 1 million logging sampling points per port 2 memory buffers per port
    Fast measurement (100 kHz) with high resolution
  • High dynamic range
    - 80 to + 10 dBm (MPM-211 / 212)
    - 70 to + 5 dBmA (MPM-215)
  • High dynamic range (logging mode)
    50 dB per scan (MPM-211 / 212)
    70 dB per scan (MPM-215)
  • Current meter module (MPM-213)
    Dynamic current range - 70 to + 10 dBmA
  • Analog output (MPM-212)
  • Up to 20-port measurement
  • Wavelength Range 1250 to 1650 nm



Applications



Specifications

Main frame, MPM-210H
Parameter
Unit
Specifications
Notes
Module number
Up to 5
 
Interface For Power meter
GP-IB, Ethernet, RS-232C
 
For System
USB
 
Trigger input
TTL (3.3 V)
BNC
Trigger output
TTL (3.3 V)
BNC
Power monitor
V
0 to 3
BNC
Supply voltage
V
AC 100 to 240, 50 / 60 Hz
 
Maximum power consumption
VA
50
 
Operating temperature
10 to 40
 
Operating humidity
%
< 80
non condensing
Weight
kg
6
 
Dimensions (W) x (D) x (H)
mm
210 x 350 x 133
 

Power meter module, MPM-211(4ch), MPM-212(2ch), MPM-215(4ch)
Parameter
Unit
MPM-211
MPM-212
MPM-215 *1
Notes
Sensor element
-
InGaAs
 
Wavelength range
nm
1250 to 1680
 
Specification wavelength range
nm
1250 to 1630
 
Power dynamic range
dBm
- 80 to + 10
- 70 to + 5  
Dynamic range @Fixed gain (typ.) *2, 3
dB
45
70 Logging mode
Number of gain range
-
5
1  
Maximum safe power
dBm
+ 16
 
Total uncertainty
%
+/- 5 @- 60 to 9 dBm
+/- 5 @>-55 dBm  
Power resolution
dB
0.001
 
 
Linearity *2, 4
dB
+/- 0.03 @- 55 to 9 dBm
+/- 0.02 @>-40 dBm
+/- 0.05 @>-50 dBm
 
Polarization dependent responsivity (typ.) *2, 5
dB
< 0.025
 
 
@1525 to 1585 nm
< 0.03
 
 
@1270 to 1630 nm
Averaging time
sec
10μ to 10
 
 
 
Data logging capability
1,000,000 x 2 buffer per port
 
 
Port number per module
4 ports
2 ports
4 ports  
Analog output *6  
None
With
None  
Connector type
FC
 
 
Current meter module, MPM-213
Parameter
Unit
Specifications
Notes
Current dynamic range
dBmA
- 70 to + 100@ 100pA to 10 mA
 
Dynamic range @Fixed gain (typ.) *2, 3
dB
45 dB
 Logging mode
Number of gain range
-
4
 
Maximum safe power dBmA 16  
Total uncertainty
%
+/- 5 @- 45 to 9 dBm
Averaging time > 10 ms
+/- 1 @>- 35 dBm (typ.)
Averaging time > 10 ms
Power resolution
dB
0.001
 
Linearity
dB
+/- 0.03@ -45 to 9 dBmA
Averaging time > 10 ms
Averaging time
sec
10μ to 10
 
Data logging capability
4 ports
 
Port number per module
1,000,000 per port x2
 
Connector type
BNC Connecter
 
Reverse Bias Voltage
V
-–
 

*1: This module is not compatible with other modules on a same Main Frame.*2: Temperature: 23 +/- 5 ℃, *3: Averaging time >50us, Within ±1 ºC after zeroing, At gain 1, 2 or 3 for MPM-211/212 *4: For MPM-211/212, Averaging time >100ms, Auto range mode, For MPM-215, Averaging time >30us *5: SMF, Straight connector
*6: Output voltage 0 to 2 V, Measurement dynamic range > 30 dB (typ.)


Download

- Brochure -

MPM-210,211,212,213 (English)

- Operation Manual -

If you require an operation manual, please download from here.

- Control Software -

If you require GUI software and drivers, please download from here.




Inquiry


Contact Us

Support / Download